About Respiration, Ventilation and Anesthesia

View Our Respiratory Products


About Rat and Mouse Ventilation

In the respiratory cycle, the inspiratory portion is typically one third of the cycle and the expiratory portion is two-thirds of the cycle. Each value can be independently adjusted to meet your research needs.

Click here for a quick link to our Tidal Volume Calculator.

Inspiratory

Pressure-controlled rat and mouse ventilators deliver a constant flow of inspiratory gas at the onset of the inspiration portion of the respiratory cycle. Gas delivery continues until reaching the inspiratory pressure threshold. Pressure in the airways and lungs is now maintained until the end of the inspiratory cycle. At this time, there is no gas flow (inspiratory pause).

Volume-controlled ventilation maintains constant inspiratory volume and allows the inspiratory pressure to vary dynamically with the tidal volume of the subject. Volume-controlled mouse ventilators deliver inspiratory gases at a constant flow for the entire period of the inspiratory portion of the respiratory cycle. The final airway pressure will be determined by the tidal volume of the subject and the connecting tubing. Care is required to avoid over pressurizing the lungs with an excessive volume of gas.

Expiratory

In the expiratory portion of the respiratory cycle, gases in the lungs are released through the expiratory tubing. The lungs are now at atmospheric pressure. During certain procedures, Peak End Expiratory Pressure (PEEP) is required to prevent the lungs from collapsing. PEEP is easily accomplished by venting the expiratory tubing through a column of water. The end of the tubing is placed under the surface of the water, at the required distance, to generate the desired PEEP.

Kent Scientific offers mouse and rat ventilators including:

  • The RoVent® module, a volume- and pressure-controlled ventilator for mice and rats up to 500 grams can be added to your PhysioSuite® physiological monitoring system. MouseVent also includes additional features such as Auto Sigh, Auto PEEP, and Auto Respiratory Assist mode.
  • The RoVent® Jr. small animal ventilator is ideal for animals up to 1,250 grams. It is fully automatic, touchscreen, has safety alarms, volume control and pressure control. 

Your specific application will determine which type of rat and mouse ventilator is best suited for your research.


About Anesthetics and Anesthesia for Small Animals

Anesthesia is the controllable and reversible loss of consciousness induced by chemical intoxication of the central nervous system. The goal of anesthetic administration is to prevent the perception of painful stimuli without undue depression of physiological functions.

Characteristic features of the anesthetic state include lowered sensitivity to outside stimuli (including pain), relaxation, and diminished motor response.

Inhalant Anesthetics

In general, the safest anesthetics are inhalants if used with the proper delivery and scavenging systems. The most commonly used, currently available inhalant anesthetics are isoflurane and sevoflurane. Sevoflurane and isoflurane both have applications in animal research due to their rapid onset and offset with minimal side effects.

  • Isoflurane: (Forane, 1-chloro-2,2,2-trifluoroethyldifluoromethyl ether) Halogenated ether that is clear, colorless, volatile liquid at standard temperature and pressure. It has a mild, ether-like odor and a molecular weight of 184.5.
  • Sevoflurane: (2,2,2-trifluoro-1-(trifluoromethyl) ethyl ether) Halogenated ether that is a sweet smelling, non-flammable fully fluorinated methyl isopropyl ether and has a molecular weight of 200.

Isoflurane and sevoflurane are the currently recommended inhalant anesthetics. Both of these agents require a precision vaporizer for delivery to the animal. Induction in mice or rats with these agents must be conducted in an induction chamber, where concentration is restricted to 5 percent. These are very rapid-acting inhalant anesthetics, and death can result if the animal is not observed closely or the concentration is too high. An advantage of these agents is that they are not metabolized and therefore have little or no toxic effects. Also, they are relatively insoluble in blood, and therefore are “blown-off” quickly, providing a quick recovery. While very similar, isoflurane and sevoflurane have slightly different effects and mechanisms of action, even though the anesthetic result is nearly identical.

Anesthesia Administration Systems

For safety, waste gas scavenging systems are required when using these agents. Precision vaporizers provide precise, controlled levels of anesthesia administered to the patient, providing a margin of safety to the animal. Using gas anesthetics at full concentration out of the bottle quickly results in overdosing the animal, and can kill very quickly.

Precision vaporizers are part of an anesthesia machine and serve to mix oxygen with the anesthetic in a precisely-controlled concentration. Animal researchers who routinely incorporate surgery or the use of anesthetics in their animal experiments should consider purchasing an anesthesia machine.

Kent Scientific carries:

  • Single and multiple channel complete anesthesia systems
  • Our new VetFlo traditional vaporizers and digital vaporizer, the SomnoSuite®, a safe, accurate gas delivery system for mice and rats.

 